MODEL 7301-800-257-3872 TOLL FREE1-978-264-0292 [email protected] EMAILwww.setra.com WEB CE and RoHS CompliantOPERATING INSTRUCTIONS
4.0 OPERATION For most accurate pressure measurement, allow the Model 730 to warm up for at least 15 minutes. After installation, periodically check
85.0 Calibration & Adjustment5. 1 Checking & Zero AdjustmentAfter installation on a system, the Model 730 should be zeroed. Figure 7 shows th
96.0 Maintenance & TroubleshootingThere are no general maintenance requirements for the Model 730 other than periodic zero adjustment. If the unit
7.0 SpecificationsPerformance Data: Accuracy 1 ± 0.5% of Reading Optional ± 0.25% of Reading Resolution: Infinite. lmite
118.0 Returning the Model 730 for Repair Please contact a Setra application engineer (800-257-3872, 978-263-1400) before returning unit for repair to
12 Calibration ServicesSetra maintains a complete calibration facility that is traceable to the National Institute of Standards & Technology (NIS
159 Swanson Road, Boxborough, MA 01719-1304 Tel: 800-257-3872/978-263-1400, Fax 978-264-0292 Email: [email protected], Web: www.setra.comSS730 Rev.
IndustrialHVACTest & MeasurementBarometricUltra High Purity/SanitaryVacuumSetra offers a complete line ofproducts for these industries:
Contents1.0 Introduction ...12.0 Mechanical Installation
1.0 Introduction Setra’s Model 730 capacitance manometer is a temperature compensated, absolute pressure transducer designed for accurate and repeata
immediately, then replace the protective flange cover and store in an area where the temperature range is controlled between -20 to +85°C. The Model
SpanZero3in.mmFigure 2: Outline Drawing of Model 730 with 5-pin Terminal Strip, and 0.5” and 0.25” OD Tube. 1.50383.0577Mating ConnectorPin IinmmModel
4Figure 3: Outline Drawing of Model 730 with 6-inch cable and15-pin D-sub Terminal Strip, 0.5” and 0.25” OD Tube.Pin 1Pin 15Model 730Model 730Model 73
5 3.0 Electrical Installation The model 730 operates from a 12-30 VDC regulated power supply for the 0-10 VDC output version or from a 9-30 VDC regu
Pin 10.123.051.5038.10Pin 9SpanZeroin.mm Pin Location Function 1 +Signal Output 9 Power Supply Common 4 Power Supply + VDC 8 Signal Output C