Installation Guide
MODEL 550
Low Range Pressure Measuring Transducers
560550-0092
Rev A
1-800-257-3872 TOLL FREE
1-978-264-0292 FAX
www.Setra.com Web Site
Zobrazit stránku 0
1 2 3 4 5 6 7 8 9

Shrnutí obsahu

Strany 1 - Installation Guide

Installation GuideMODEL 550Low Range Pressure Measuring Transducers560550-0092Rev A1-800-257-3872 TOLL FREE1-978-264-0292 FAXwww.Setra.com Web Site

Strany 2 - Ultra High Purity / Sanitary

Setra offers a complete line ofproducts for these industries:IndustrialHVACTest & MeasurementBarometricUltra High Purity / Sanitary

Strany 3 - ACTION ON RECEIPT

Please Note: Ranges above 7 psi are designed and manufactured in accordance with soundengineering practice as defined by the Pressure Equipment Dire

Strany 4 - MECHANICAL INSTALLATION

GENERAL* Transducer should not be subjected to greater than the maximum allowable pressure ortemperature as defined in the transducer specification.*

Strany 5

ELECTRICAL INSTALLATIONAll types include suppression devices providing transient protection to EN 61000-4-2 and EN 61000-4-4.For all types conformity

Strany 6 - W 110 to 74%

To maintain sealing integrity ensure end cap ‘O’ ring is correctly seated, and clean prior to re-fitting.RANGE SWITCH DETAILSSwitch Position(Current M

Strany 7 - RETURN TO FACTORY

MAINTENANCERoutine Inspection: Not required except for periodic inspection of the cable and molding to ensurethat these are neither damaged nor softe

Strany 8 - W A R N I N G

W A R N I N G * IF INSTRUMENT IS SUBJECTED TO +400°C PLUS Some instruments use Viton sealant. Above +400°C this material decomposes produ