CHANGING THE BAUD RATE AND PARITY
The baud rate and parity of the DATUM 2000™ Serial Option can be set to
the following values:
COMMAND ASCII
BAUD: 300 3 33H
600 6 36H
1200 1 31H
2400 2 32H
4800 4 34H
9600 9 39H
PARITY: NONE N 4EH
ODD O 4FH
EVEN E 45H
To change the baud or parity, transmit a P, ASCII 50H, and the command
number or letter from the table above. For example, to set the DATUM
2000™ to 9600 baud and even parity, transmit P (always allow 1 second
between commands sent to DATUM 2000™) and then transmit 9, to set the
baud rate to 9600. Set your computer or terminal to 9600 baud and then
transmit P and E to set even parity.
When you change the communications parameters, they are stored internally
and will be recalled every time the DATUM 2000™ is powered up.
Hint: Transmit a V to the DATUM 2000, if the verify message "SETRA
DATUM 2000", is not received, wait 1 second, and retransmit a V. Transmit
a V each second until the verification message is received. When the
verification message is received, the DATUM 2000™ is ready for a new
command. Now transmit a P, wait at least 1 second and transmit the baud or
parity parameter.